| Dosimeters: for the measurement of radiation dose in industrial irradiation plants | ||||
| Type | ||||
| Material a single batch of "Perspex" cast polymethylmethacrylate sheets |
. | . | . | |
| Dimensions length 30 mm; width 11 mm |
3 +/- 0.55mm | 3 +/- 0.55mm | 1.7mm nominal | |
| Measurement reproducibility Coefficient of variation of specific absorbance measurements on sets of dosimeters from the batch, simultaneously irradiated together, in a radiation field uniform within +/- 1%, is, over the entire calibrated range: |
not more than 2%
5 to 50 kGy |
not more than 2.5%
1 to 10 kGy at 603nm wavelength 10 to 30kGy at 651nm wavelength |
not more than 3%
100 to 3000Gy |
|
| Calibration quality Over the calibrated range, all mean specific absorbance data points are within 3% of a fourth order polynomial least squares fit to the data. |
. | . | . | |
| Traceability HARWELL calibration is traceable to the UK standard of absorbed dose at the National Physical Laboratory |
. | . | . | |
| Shelf life Shelf life demonstrated to be at least |
ten years | five years | . | |
| Packaging Sealed in labelled sachets made of polyester / aluminium foil / polyester / polyethylene laminate packed 1000 per carton Packing is to defined AQLs, checked to ISO 2859-1, single normal inspection, level II. |
. | . | . | |
| Typical response curves(click to view) | ||||