Return to home page

Dosimeter
Specifications

Return to home page


HARWELL Dosimeters Ltd
540 Becquerel Avenue
Harwell Science and Innovation Campus
DIDCOT, Oxfordshire
OX11 0RA
United Kingdom
Tel: +44 (0)1235 435704
Fax:+44 (0)1235 436313
E-mail Harwell Dosimeters


blue_bar.gif (71 bytes)


© 2008 Harwell Dosimeters Ltd

Dosimeters: for the measurement of radiation dose in industrial irradiation plants

Type

View response curve

View response curve

View response curve


Material
a single batch of "Perspex" cast polymethylmethacrylate sheets
. . .

Dimensions
length 30 mm; width 11 mm
3 +/- 0.55mm 3 +/- 0.55mm 1.7mm nominal

Measurement reproducibility
Coefficient of variation of specific absorbance measurements on sets of dosimeters from the batch, simultaneously irradiated together, in a radiation field uniform within +/- 1%, is, over the entire calibrated range:
not more than 2%

5 to 50 kGy

not more than 2.5%

1 to 10 kGy at 603nm wavelength

10 to 30kGy at 651nm wavelength

not more than 3%

100 to 3000Gy


Calibration quality
Over the calibrated range, all mean specific absorbance data points are within 3% of a fourth order polynomial least squares fit to the data.
. . .

Traceability
HARWELL calibration is traceable to the UK standard of absorbed dose at the National Physical Laboratory
. . .

Shelf life
Shelf life demonstrated to be at least
ten years five years .

Packaging
Sealed in labelled sachets made of polyester / aluminium foil / polyester / polyethylene laminate packed 1000 per carton

Packing is to defined AQLs, checked to ISO 2859-1, single normal inspection, level II.

. . .

Typical response curves (click to view)

View response curve

View response curve

View response curve